The phantom for small FOV NM and PET systems performance evaluation (collimator, artifacts, calibration, reconstruction parameters) Among other things, it can be used to evaluate: center-of-rotation error, non-uniformity artifacts, changes of radius-of-rotation on spatial resolution, reconstruction filters on spatial resolution, attenuation and scatter compensation.
- complies with:
- NEMA Standards Publication (NU 1-2001) Performance Measurements of Scintillation Cameras
- AAPM Report No. 9 - Computer Aided Scintillation Camera Acceptance Testing
- AAPM Report No. 22 - Rotating Scintillation Camera SPECT Acceptance Testing and Quality Control
- ACR–SNM (Res. 5 – 2011) technical standard for diagnostic procedures using radiopharmaceuticals
- CE certified
- the manual provides detailed guidelines for carrying out each test, results assessment and registration
- main cylinder:
- inside cylinder diameter: 140 mm
- inside cylinder height: 150 mm
- cylinder wall thickness: 6 mm
- cold rods insert:
- rod diameters: 3.2, 4.8, 6.4, 7.9, 9.5 and 11.1 mm – height of rods: 40 mm
- cold spheres:
- solid sphere diameters: 3.2, 4.8, 9.5, 12.7, 15.9 and 19.1 mm – height of the center of the spheres from the base plate: 78.5 mm
- optional PET Lid with cylindrical samples:
- refillable thin-walled cylinders, diameters: 4, 6, 8, 12, 16 mm
- PTFE solid cylinder diameter: 16 mm
- cylinder height: 25.4 mm
- optional heavy duty carrying case (08-120)
Documents
Sorry, there are no documents available for this product.