The phantom can test the performance of a mammographic system and its ability to image small structures similar to those found clinically: calcifications, fibrous calcifications in ducts and tumour masses.
Product features:
- ACR Accredited
- complies with:
– IEC 61223-3-2
– ACR Mammography Accreditation Program Requirements (May 4, 2012) - CE certified
- the manual provides detailed guidelines for carrying out each test, results assessment and registration
Technical data (can be modified to customer specifications):
- dimensions: 101.5 x 108.0 x 44.0 mm
- simulates 42 mm compressed breast of average glandular/adipose composition
- total PMMA thickness: 36.75 mm
- wax insert:
– thickness: 7.25 mm
– nylon fibrils diameters: 1.56, 1.12, 0.89, 0.75, 0.54 and 0.40 mm
– microcalcifications: 0.54, 0.40, 0.32, 0.24 and 0.16 mm Al₂O₃ specks
– tumor-like masses: 2.00, l.00, 0.75, 0.50 and 0.25 mm thick - 4 mm thick PMMA contrast disc
- comfortable carrying case
Documents
Sorry, there are no documents available for this product.