For radiopharmaceutical quality control
The Aluminium Breakthru Kit provides a rapid, easy and inexpensive way to test aqueous solutions, particularly pertechnetate generator eluate, for trace quantities of aluminium. Aluminium forms an intense red precipitate with the indicator paper, and the intensity of the colour is directly proportional to the amount of aluminium in the solution. The British/European Pharmacopeia allows a concentration of <5 micrograms per millilitre (5 µg ml) in technetium 99m eluate from a fission molybdenum generator.
Simple Procedure:
- Place a drop of the eluate or solution to be tested on the indicator paper. The best procedure is to form a hanging drop using a 19-22G needle.
- Place a drop of the standard aluminium solution on the indicator paper. Use the same size drop.
- Compare the intensity of the red spot formed. If the eluate spot is less intense than the standard solution, the eluate contains less than 10 µg/ml aluminium.
Models
150-780 | Chromatography Kit, Tec-Control Aluminium Breakthru Kit includes: Aluminium standard, 5 ml, 10 µg/ml, 50 indicator strips and manual |
150-785 | Chromatography Kit, Tec-Control Aluminium Breakthru Kit includes: Aluminium standard, 5 ml, 5 µg/ml, 50 indicator strips and manual |
Solvents and Strips
Radiopharmaceutical | Solvents(s) required | Strips required |
Aluminium Breakthru | 150-781 10 ug/ml | 150-782 |
Aluminium Breakthru | 150-783 5 ug/ml | 150-782 |
Documents
Sorry, there are no documents available for this product.