The phantom for performance evaluation of small animal systems or ultra-high resolution ECT systems. It can be used to evaluate: center-of-rotation error, non-uniformity artifacts and spatial resolution. Available in two sizes for FOV at least 45 mm or 75 mm.

  • complies with:
    • NEMA Standards Publication (NU 1-2001) Performance Measurements of Scintillation Cameras
    • AAPM Report No. 9 - Computer Aided Scintillation Camera Acceptance Testing
    • AAPM Report No. 22 - Rotating Scintillation Camera SPECT Acceptance Testing and Quality Control
    • ACR–SNM (Res. 5 – 2011) technical standard for diagnostic procedures using radiopharmaceuticals
  • CE certified
  • the manual provides detailed guidelines for carrying out each test, results assessment and registration

at least 75 mm FOV version:

  • main cylinder:
    • outside cylinder diameter: 80 mm
    • inside cylinder diameter: 74 mm
    • inside cylinder height: 65 mm
  • cold rods insert:
    • rod diameters: 1.2, 1.6, 2.4, 3.2, 4.0 and 4.8 mm
    • height of rods: 34 mm

At least 45 mm FOV version:

  • main cylinder:
    • outside cylinder diameter: 50 mm
    • inside cylinder diameter: 45 mm
    • inside cylinder height: 65 mm
  • cold rods insert:
    • rod diameters: 1.2, 1.6, 2.4, 3.2, 4.0 and 4.8 mm
    • height of rods: 34 mm
  • optional carrying case