Radiation/Light Field Analyzer
The Visi-X measures the alignment between radiation and light field and is a proven concept in Quality Control and Service. It is a cassette shaped instrument for checking the light and radiation field coincidence for X-ray equipment. The Visi-X can also be used for checking of the centering of the bucky tray.
Instant Visualization
The Visi-X is based on a after-glowing phosphorus screen. Simply darken the X-ray room and place your Visi-X under the X-ray tube. Adjust the light field and make your exposure. The radiation field will immediately be visualized by the glow of the special phosphor compound. Misalignment down to ±1 mm will be clearly shown on the built-in scales (afterglow will last for several minutes.) No film is needed; therefore, no time is lost going back and forth to the film developer, if there is one.
Protective Daylight Filter
The phosphor is non-radioactive and is covered by perspex plates. A daylight filter protects the phosphor from accidental excitation from light sources. The lifetime expectancy of the phosphor is not affected by light or X-rays having an energy within the recommended range.
Detection phosphor Afterglowing rare-earth | |
---|---|
Emission Color | Green |
Operating Temperature | +15°C to +45°C |
Dimensions | 320 x 276 x 11 mm (without daylight filter) |
Equivalent Cassette Size | 24 x 30 cm |
Weight | 1.4 kg |
Other | Daylight filter, magnetic lock, ruler, documentation chart |
Option | Carrying case |
Field Position Anal | yser | |
---|---|---|
Field Positioning Accuracy | < ±0.5 mm | |
Centering Accuracy | < ±0.5 mm | |
Scale Range | Circular Fields | 5-6 cm diameter |
Square Fields | 5 x 5, 10 x 10, 15 x 15 and 20 x 20 cm | |
Indicated Deviation | ±10 mm | |
Scale Accuracy | ±0.1 mm | |
Recommended Output | 130 µGy/mAs at 100 kVp and 75 cm S.I.D. | |
Usable Energy Range | 15 - 200 keV |
Documents
-
Visi-X
Download (798.72 kB)
Catalogues
-
Radiology-Quality-Assurance-EU-Version-1.4
Download (18.94 MB)