MRI Phantom

ACR accredited Medium MRI phantom for comprehensive evaluation of critical imaging parameters of magnetic resonance imaging (MRI) in a time-efficient manner. The phantom can be used for the measurement of absolute values for calibration purposes. However, its design is optimized for quick daily Quality Assurance too. 

The Pro-MRI ACR Medium can be used to measure:

  • geometric distortion
  • spatial resolution
  • slice thickness and position
  • interslice Gap
  • image bandwidth
  • low contrast detectability
  • image uniformity
  • signal-to-noise ratio (SNR)
  • physical and electronic slice offset
  • point of reference
  • bandwidth: water-fat shift

Features

  • ACR accredited
  • Complies with:
    • ACR MRI Accreditation Program
    • IPEM Report 80 “Quality Control in MRI”, 1998
    • AAPM Rep. 28 “Quality Assurance methods & phantoms for MRI”, 1990
    • AAPM Rep. 34 “Acceptance testing of MRI systems”, 1992
    • AAPM Rep. 100 “Acceptance & Quality Assurance procedures for MRI facilities”, 2010
  • CE certified

Basic Specifications

  • outside cylinder diameter: 178 mm
  • outside cylinder height: 157 mm
  • inside cylinder diameter: 165 mm
  • inside cylinder height: 134 mm
  • the phantom is filled with 10 millimolar (mmol) nickel chloride solution containing 0.45% by weight sodium chloride
  • the outside of each phantom has the words “NOSE” and “CHIN” etched into it as an aid to orienting the phantom for scanning, as if it were a head

Resolution insert

  • four matrices of holes, diameters: 0.8, 0.9, 1.0 and 1.1 mm
  • spaces between the holes are equal to hole diameters
  • two small containers with water and fat for water-fat shift evaluation

Slice thickness insert

  • 1 mm wide counter-descending slits on both sides
  • slits form two ramps descending at 1:10

Geometric distortion insert

  • 3x3 holes array, 40 mm spacing,
  • 2 mm diameter
  • 120 mm on a side
  • 10 mm thick

Low contrast insert

  • partial volume contribution of these sheets and filling solution produce contrasts: 1.4, 2.5, 3.6, and 5.1%
  • each disc contains 10 groups of 3 holes arranged in spokes
  • each spoke has the same diameter
  • diameters range from 7.0 to 1.5 mm

Slice position accuracy

  • one set of paired 45° wedges. The distance between the intersection of the wedges is 100 mm

Product included

  • hard-sided carrying case (MR unsafe)

Documents

Sorry, there are no documents available for this product.

The LabLogic Group will always treat your details with the utmost care and will never sell them to third party companies.

Please choose one of the following:



Previous
Next