This scatter phantom simulates in-vivo forward and backscatter characteristics of 99mTc gamma rays for the extrinsic measurement of a scintillation camera’s deadtime. The phantom produces a spectrum typical of that observed from 99mTc in the myocardium. Reference: Ralph Adams, Gerald J. Hine, and C. Duane Zimmerman, “Deadtime Measurements in Scintillation Cameras Under Scatter Conditions Simulating Quantitative Nuclear Cardiography,” The Journal of Nuclear Medicine, 19 (1978), 538-544.

  • Complies with:
    • NEMA Standards Publication (NU 1-2001) Performance Measurements of Scintillation Cameras
    • NEMA Standards Publication (NU-1 2007) Gamma Cameras
    • AAPM Report No. 9 - Computer Aided Scintillation Camera Acceptance Testing
    • AAPM Report No. 22 - Rotating Scintillation Camera SPECT Acceptance Testing and Quality Control
    • ACR–SNM (Res. 5 – 2011) technical standard for diagnostic procedures using radiopharmaceuticals
  • CE certified
  • The manual provides detailed guidelines for carrying out each test, results assessment and registration
  • made of PMMA
  • dimensions: 150 x 200 x 200 mm
  • the two holes are used to hold the radioactive sources
    • hole dimensions: Ø 17 mm x 120 mm deep
    • spaced 50 mm apart (center-to-center)
    • distance from the face of the phantom: 50 mm

Documents

Sorry, there are no documents available for this product.

The LabLogic Group will always treat your details with the utmost care and will never sell them to third party companies.

Please choose one of the following:



    Previous
    Next