The phantom for high-resolution NM and PET systems performance evaluation (collimator, artifacts, calibration, reconstruction parameters). Among other things, it can be used to evaluate: center-of-rotation error, non-uniformity artifacts, changes of radius-of-rotation on spatial resolution, reconstruction filters on spatial resolution, attenuation and scatter compensation.

  • complies with:
    • NEMA Standards Publication (NU 1-2001) Performance Measurements of Scintillation Cameras
    • AAPM Report No. 9 - Computer Aided Scintillation Camera Acceptance Testing
    • AAPM Report No. 22 - Rotating Scintillation Camera SPECT Acceptance Testing and Quality Control
    • ACR–SNM (Res. 5 – 2011) technical standard for diagnostic procedures using radiopharmaceuticals
  • CE certified
  • the manual provides detailed guidelines for carrying out each test, results assessment and registration
  • main cylinder:
    • inside cylinder diameter: 206 mm
    • inside cylinder height: 186 mm
    • cylinder wall thickness: 7 mm
  • cold rods insert:
    • rod diameters: 3.2, 4.8, 6.4, 7.9, 9.5 and 11.1 mm
    • height of rods: 88 mm
  • cold spheres:
    • solid sphere diameters: 9.5, 12.7, 15.9, 19.1, 25.4 and 31.8 mm
    • height of the center of the spheres from the base plate: 127 mm
  • optional PET Lid with cylindrical samples:
    • refillable thin-walled cylinders, diameters: 8, 12, 16 and 25 mm
    • water-filled cylinder diameter: 25 mm
    • air-filled cylinder diameter: 25 mm
    • PTFE solid cylinder diameter: 25 mm
    • cylinder height: 38 mm
  • optional heavy-duty carrying case

Documents

Sorry, there are no documents available for this product.

The LabLogic Group will always treat your details with the utmost care and will never sell them to third party companies.

Please choose one of the following:



Previous
Next