Ionization Chamber for CTDI Measurements
The DCT10 is a rugged, cylindrical pencil shaped air ionization chamber for CTDI measurements. The CT chamber is intended for measuring and monitoring the exposure output level of CT scanners in a phantom or in air. CT dose index, CTDI, can be measured in accordance with IEC 61223-2-6.
The DCT10 can be connected via Chamber Adapter to the Piranha or Cobia to measure CTDI.
The DCT10 RS can be connected directly to the Barracuda (via an electrometer module supporting bias) to measure CTDI.
Ranges with Piranha & Cobia Flex (Chamber Adapter required)
Dose Length Rate | 0.8 mGycm/s to 8 Gycm/s |
Inaccuracy | ±6 % or ±0.08 mGycm/s |
Connector type | LEMO triaxial |
Cable | 2 m, low noise triaxial |
Active volume | 4.9 cm3 |
Active length | 100 mm |
Inner diameter | 8.0 mm |
Typical leakage | ±4 fA |
Radiation quality | 80 - 150 kV |
Sensitivity | 400 nC/Gy or 3.5 nC/R |
Documents
-
Accessory Brochure
Download (1.89 MB)
Catalogues
-
Radiology-Quality-Assurance-EU-Version-1.4
Download (18.94 MB)