Alpha, Beta and Gamma contamination probe, powered by the 6150AD5/6
- End window GM
- Suitable for low intensity X-ray or gamma radiation
Analog range: 0.1 s¯¹ - 10k s¯¹
Digital range: 0.01 s¯¹ - 9.99k s¯¹
Detector | End window tube LND 7231 or equivalent, not energy compensated, gamma sensitivity at 137Cs approx. 5600 pulses per µSv |
Tube window or tube wall resectively | Dimensions: Diameter 28mm that is 6.2 cm2 in areaMaterial: Mica, areal density 1.5 - 2 mg/cm2 |
Range | 0.01 S-1 to 10 kS-1 |
Temperature range | -30°C to +50°C (137Cs gamma radiation free in air) |
Humidity and pressure | 0 - 95% relative humidityAtmospheric pressure 60 - 130 kPa (600 - 1300 mbar) |
Power Supply | 4.75 V out of 6150AD |
Dimensions | 40 (diameter) x 132 (length) mm |
Weight | ~ 180 g including protective cap |
Probe cable | Max. 100 m |
Documents
-
SSL AD1719k
Download (789.7 kB)
Catalogues
-
Monitoring-and-Analytical-Equipment-EU-Version-1.3
Download (28.21 MB)